Operation and Maintenance of Low-Speed Refrigerated Centrifuges
Release time:
2012-01-04
Overview:
The LUXIANGYI DDL5M low-speed refrigerated centrifuge, in addition to the typical structural and functional features of conventional centrifuges, is equipped with systems for refrigeration, vibration reduction, and instrument control. Clinically, it is primarily used for separating, precipitating, and concentrating sample solutions to extract components of different densities. Therefore, proper operation and reasonable maintenance are crucial for ensuring the instrument’s normal functioning and extending its service life.
Instrument Operation Precautions:
1. To prevent contamination of the instrument by radioactive isotopes, first carefully inspect the sample tubes for cracks or pinholes to avoid leakage during centrifugation. Second, handle the plastic tubes placed in the centrifuge cups with great care to prevent liquid from splashing. If liquid enters the centrifuge chamber and causes contamination, immediately remove the centrifuge cup, rinse it thoroughly several times with tap water, and allow it to air dry. At the same time, perform a comprehensive cleaning of the entire centrifuge chamber.
2. Maintain a stable operating environment both inside and outside the centrifuge. The DDL-5M centrifuge should be placed horizontally, with a clearance of at least 10 cm from the wall, and in a well-ventilated area. It should be kept away from direct exposure to heat sources and sunlight. The room temperature should not exceed 30°C; otherwise, the cooling performance may be affected. When the ambient humidity is high, moisture in the centrifuge chamber will evaporate more slowly. In such cases, promptly remove any accumulated moisture to prevent rusting of the instrument. When centrifuging sample tubes just removed from the water bath, frequently check whether there is any accumulated water in the centrifuge rotor and promptly empty it to keep the rotor dry and ensure balance during centrifugation.
3. To prevent aging of the sealing ring, a rubber sealing ring is installed at the contact point between the upper part of the centrifuge’s inner chamber and the lid door. As usage time increases, you should regularly check for any deformation to avoid external heat infiltration, which could impair the cooling effect and affect the accuracy of test results.
4. Set the temperature correctly. When setting the temperature, the upper limit must be higher than the lower limit; otherwise, the temperature set for the lower limit will be 2℃ lower than that set for the upper limit.
5. Set the rotational speed correctly. When setting the rotational speed, you must do so only after the timer has been turned on. First, convert the desired rotational speed into frequency (1 Hz = 60 rpm). Then, press the △ or v key to adjust the frequency to the required value. After a few seconds, the frequency will automatically be converted into the corresponding rotational speed. While the instrument is running and the rotational speed is being changed, do not use the SHIUF key; otherwise, it may cause the program to malfunction. Be sure not to set the rotational speed beyond the maximum allowable limit to ensure the safe operation of the instrument.
6. If the machine displays “0.130” or any other abnormal reading during operation, turn off the power and unplug the machine for 10 seconds before restarting it. After the set speed is displayed, press the reverse button.
Maintenance and upkeep:
1. Regular maintenance: As with the maintenance of other medical instruments, in addition to routine inspections and repairs of electrical circuits, electronic components, and mechanical moving parts, pay attention to whether various indicator lights are damaged or aged.
2. Ensure that the centrifuge tubes are symmetrically balanced. Although the centrifuge is equipped with a secondary vibration-damping system, if the sample tubes placed inside the centrifuge chamber are unbalanced, it can cause the centrifuge to shake and shift. After stopping the machine and resetting it, carefully check whether the instrument is functioning properly.
3. The latch on the centrifuge lid should be checked regularly to ensure it remains flexible and prevents forceful opening or closing. If any signs of poor contact are detected, the latch should be promptly repaired or replaced. For common faults, pay close attention and observe them carefully on a regular basis, and address them promptly.
4. Properly record the operating status of the centrifuge and the duration of each run. Document the causes of instrument malfunctions, the troubleshooting methods used, and the times when these issues were resolved, ensuring that the instrument operates in an optimal condition.