How to Troubleshoot Unstable Rotational Speed in Low-Temperature Centrifuges
Release time:
2013-05-30
Overview:
Fault phenomenon: Low-temperature centrifuges are essential equipment in hospitals and research institutions, and their performance directly affects the quality of the samples being centrifuged.
Troubleshooting: The motor speed is unstable. From a theoretical perspective, the following four scenarios could all be possible.
(1) Adjustable resistors, being installed on the control panel and used very frequently, commonly experience poor contact over time.
(2) There are also quite a few problems with thyristors, typically including breakdown and short circuits, open circuits, failure to trigger, or poor internal contact.
(3) One of the four rectifier diodes has a problem.
(4) There is carbon dust between the motor rotor and the commutator, or the carbon brushes themselves are defective.
Following the repair principle of tackling easier tasks before more difficult ones, we first measured the adjustable resistor. The resistance value changed regularly as the adjustment was made, confirming that the potentiometer was in good condition. Next, we measured the thyristor. Then we proceeded to measure the four rectifier diodes. Since these diodes are high-current, gold-sealed devices, one end is secured to the circuit board with a screw, while the other end is soldered onto the board using wires. When removing them from the circuit board, we carefully checked whether the positive and negative terminals were loose. While measuring each diode, we gently shook it simultaneously to verify its integrity. Additionally, during installation, it’s crucial to use components of the same model and specifications whenever possible. Finally, after powering on the motor and eliminating the previously identified faults, the motor speed promptly returned to normal operation.