Introduction to the Safe Use of Laboratory Medical Centrifuges
Release time:
2011-12-26
Overview:
When using a centrifuge, strictly follow the manufacturer’s installation and commissioning instructions.
(1) Power supply voltage: Voltage fluctuations must comply with the national standard within ±10%; otherwise, the centrifuge will not operate properly. If voltage fluctuations exceed this range, it is recommended to use a voltage regulator to ensure compliance with the power grid requirements.
(2) The grounding wire must be secure: The wiring of the room’s power supply must meet the required standards. In China, the power system uses a three-phase four-wire configuration. The three phases refer to the three-phase electricity used in industrial settings. When using three-phase power, the neutral wire and the ground wire must be kept separate. More importantly, whether it’s three-phase industrial power or ordinary single-phase power, the ground wire must be reliably connected to prevent electric leakage. Under no circumstances should the ground wire be connected to heating pipes or water pipes.
(3) The floor must be level and stable: For large centrifuges, the floor at the installation site must be firmly secured to ensure that the centrifuge rotor operates normally without sinking or becoming uneven. The floor at the installation and fixing locations should be both level and stable.
(4) Adjust the level: Find a small, relatively accurate spirit level. With the centrifuge lid open, locate the horizontal position on the spindle. Adjust the level by turning the adjusting bolts located next to the four casters. Be sure that none of the four feet is unsupported.
The above information is provided by Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. Please cite the source website when reprinting.