High-speed benchtop centrifuge technology for separating serum lipoproteins

Release time:

2013-01-24


Overview:

Electrophoresis has proven particularly effective for separating fractions with high apolipoprotein content (HDL and VHDL), whereas centrifugation is more suitable for separating serum lipoproteins of various densities.

Serum lipoproteins are heterogeneous complexes composed of lipids and certain specific proteins found in the blood. Due to varying ratios of lipids to apolipoproteins, their density ranges from 0.96 g/mL.

or as low as between 1.21 g/mL. In paper electrophoresis, four bands are observed: chylomicrons, very-low-density lipoproteins, low-density lipoproteins, and high-density lipoproteins. Each type of lipoprotein can further be subdivided into more categories. Subcomponent.
Electrophoresis has proven particularly effective for separating fractions with high apolipoprotein content (HDL and VHDL), whereas centrifugation is more suitable for separating serum lipoproteins of various densities.

1. Gradient Materials: Serum lipoproteins have relatively low density. Commonly used gradient materials include NaCl, NaI, KBr, and KI, while the commonly employed buffer is EDTA.

2. Pre-separation:
1. Separation of blood cells from serum:

Centrifugation parameters: Whole blood, swing-out rotor, 3,000 rpm × 20 min
Result: The precipitate is blood cells, and the upper layer is serum.

2. Chylomicron Separation: In human plasma from individuals who have fasted for more than 12 hours, the chylomicron content is extremely low. For those who have recently eaten, since their plasma contains high levels of chylomicrons, chylomicrons should first be removed by centrifugation.
Centrifugation parameters: 4 × 10 (g×min), 10°C. Any rotor speed is acceptable.
Gradient solution preparation: Fill the lower 3/4 of the centrifuge tube with plasma, and the upper 1/4 with a solution containing 0.5 M NaCl plus 0.3 M EDTA, pH 7.4.
Result: The chylomicrons floated to the surface and were aspirated.

3. Serum protein separation: Remove globulins, albumin, and other proteins.
Centrifugation parameters: (2.5–3) × 10⁴ (g×hr); applicable to all types of rotors. For example, using an angle rotor at 70,000 rpm for 6 hours at 10°C.
Configuration: The tube contains 1/3 serum and 2/3 of a solution with a density of 1.31 g/ml, composed of NaCl plus NaBr; after stirring, the final density is 1.21 g/ml.
Result: The upper 1/6 of the tube contains serum lipoproteins, while the lower 5/6 contains other proteins.

3. Methods:
1. Sequential Flotation Method:
a) The ultracentrifugation method: This is the earliest method employed, involving 4,000 rpm × 24 hours × 3 cycles. The 10°C NaCl + NaBr gradient is floated in the order of 1.006, 1.063, and 1.21, with samples taken from the upper part of the tube at each step to obtain fractions of different densities.

Lipoproteins. Simple operation, high resolution, high yield, time-consuming, high cost.
b) Ultrahigh-speed method: 100,000 rpm × 2.5 hours × 2 times, 10°C; 100,000 rpm × 4 hours × 1 time; NaCl + KBr + EDTA gradient to isolate VLDL, LDL, and HDL by density-gradient centrifugation without density media.

2. Single-spin method:
a) Single-level separation
Using a rotor with a maximum speed of 25,000–28,000 rpm and a 6×40 ml swing-out rotor, prepare an 8-layer gradient ranging from 1.006 to 1.21 using solutions containing 1.006 g/ml NaCl and 1.35 g/ml (NaCl + KBr + EDTA).

A continuous gradient was prepared, with the bottom layer consisting of a sample solution composed of serum and KBr at a concentration of 1.25 g/mL. The sample was subjected to centrifugation at 25,000 rpm for 4 hours at 4°C, yielding distinct banding regions corresponding to VLDL, LDL, and HDL.
Using a 6×13 ml swing-out rotor at the maximum speed of 40,000 rpm, centrifuge at 40,000 rpm for 24 hours at 20°C with a discontinuous KBr gradient ranging from 1.006 to 1.21 g/ml. The tubes are slender, showing clear layering, and the recovery rate is satisfactory.

High, long centrifugation time
b) Single-stage vertical tube separation:
The discontinuous gradient single-tube capacity for NaCl/KBr or NaCl/NaBr ranges from 5 ml to 40 ml, with rotor speeds ranging from 50,000 rpm to 80,000 rpm. Centrifugation lasts (0.5–3) hours, and the serum is adjusted to a density of 1 using KBr or (NaBr) in the lower layer.

The upper layer has a density of 3 g/mL, while the lower layer consists of a 1.006 g/mL NaCl solution at 10°C–20°C; after one increase and deceleration step, separate the layers. In the discontinuous NaCl/sucrose gradient, the lower layer contains 48% w/w sucrose, and the middle layer comprises serum in 4 M NaCl.

In the middle, the upper layer consists of 0.67 M NaCl plus 0.05% EDTA; the centrifugation parameters are the same as above, but the duration is longer.
c) Large-capacity separation using a swing-out rotor:
Centrifugation parameters: Swing-out rotor, maximum speed 30,000–48,000 rpm; capacity ranging from 650 ml to 1,900 ml; capable of separating 50–150 ml of serum per run. Low-speed sample loading and unloading; separation speed: for large rotors.

(25,000–28,000 rpm) × 24 hours, small rotor (30,000–38,000 rpm) × 1 hour, 10°C–20°C
Gradient solution: A discontinuous or linear gradient of NaCl + NaBr or KBr + EDTA. The sample is placed at the layer of maximum density, near the rotor’s core, with pure water serving as the separation layer. The outermost region of the separation chamber is filled with a high-density NaCl/KBr solution.

NaCl/NaBr as a buffer layer.
Collection: After pumping, the sample is detected by a UV detector along the flow path, and then automatically collected by a fraction collector.