Operating and Maintenance Procedures for Medical High-Speed Centrifuges

Release time:

2015-04-01


Overview:

Our product lineup includes high-speed centrifuges, low-speed centrifuges, large-capacity centrifuges, refrigerated centrifuges, and more. We also offer custom-made centrifuges tailored to specific customer needs. Our products are widely used in research and production institutions across various fields, including agricultural science, bioengineering, food processing, chemical industry, pharmaceuticals, clinical medicine, blood banks and blood stations, inspection and quarantine, disease control and prevention, and environmental protection.

Medical high-speed centrifuges operate at high speeds, so safety must be carefully observed.

The following points must be achieved:

1. When using a centrifuge, place it on a stable and sturdy workbench. All centrifuge bases are equipped with rubber suction feet that, thanks to atmospheric pressure and the instrument’s own weight, ensure tight adhesion to the workbench. Large-capacity low-speed centrifuges and high-speed refrigerated centrifuges should be placed on a solid, level floor accordingly. Ultracentrifuges, which can weigh hundreds of kilograms, must be installed on a very solid floor and equipped with dustproof and moisture-proof devices to ensure that the centrifuge chamber maintains a sufficient vacuum for normal operation. Before starting the centrifuge, make sure the load is balanced; the smaller the weight discrepancy, the better. Otherwise, severe vibrations may occur, damaging the rotor and spindle.

2. Select an appropriate rotor and control its rotational speed. The maximum rotational speed of the rotor should be reduced under the following circumstances: when the rotor’s operating time and number of runs reach their specified service life; when the rotor exhibits localized surface damage or slight corrosion inside the tube holes; when using stainless steel centrifuge tubes, sleeves, thick-walled tubes, or centrifuge bottles; when the average density of the sample exceeds 1.2 g/cm³; or when the rotor is exposed to high temperatures.

3. Before each use, the rotor must be carefully inspected to ensure that there are no foreign objects or contaminants inside the holes, in order to maintain balance. The rotor must be kept dry and clean, and must never be subjected to impacts or scratches. If the rotor accidentally falls, it should be examined by X-ray to confirm that there is no internal damage before it can be used again. After each use, the rotor should be soaked and cleaned in warm water (50℃~60℃) containing a neutral detergent, then rinsed thoroughly with distilled water or deionized water. After drying it with a soft cloth, use a hairdryer to further dry it, apply wax, and store it in a dry place.

4. The centrifuge rotor must be operated only after the centrifuge tubes have been properly placed; running the centrifuge without any tubes loaded is strictly prohibited. When starting the centrifuge, set the rotation speed or centrifugal force to a low setting, and gradually increase it to the desired level once the machine has started. After each run, be sure to return the rotation speed and centrifugal force to the low setting before starting another run. Never attempt to start the centrifuge directly at a high rotation speed or high centrifugal force setting.

5. Centrifuge tubes of various types differ in tensile strength and elongation; they should be selected according to specific requirements. Centrifuge tubes can be made from many different materials, such as polyethylene, cellulose, polycarbonate, and polypropylene. When using these tubes, ensure that the samples, solvents, and gradient materials do not corrode the centrifuge tubes or their cap components. Avoid using centrifuge tubes that are expired, aged, cracked, or already corroded, or reduce their operating speed if used under such conditions. Monitor and limit the number of times plastic centrifuge tubes are used, and pay attention to matching them with the appropriate specifications. During centrifugation, never open the chamber lid, and never allow your hands or foreign objects to come into contact with the rotor or centrifuge tubes while they are rotating.

6. For low-temperature centrifugation samples, first pre-cool an empty rotor at 2000 rpm for a certain period of time. During pre-cooling, maintain the temperature around 0℃. Alternatively, the rotor can be placed in a refrigerator for several hours to pre-cool and kept ready for use. Centrifuge tubes can be directly stored in the refrigerator for pre-cooling.

7. During the centrifugation process, if any abnormal phenomena are detected—such as unusual noises or vibrations—the machine should be stopped immediately for inspection. Do not resume operation until the cause has been identified.

8. During daily use of the centrifuge, the horizontal alignment of the main unit should be checked and adjusted every three months. The vacuum pump oil should be replaced after every 500 million revolutions of operation. Every approximately 1,500 hours of use, the bearings in the drive section should be cleaned and lubricated with high-speed grease. The joint between the spindle and the rotor should be regularly greased to prevent rusting. When the centrifuge is not used for an extended period, it should be coated with anti-rust oil and wrapped in oil-resistant paper. During periods of non-use, the centrifuge should be started at low speed once or twice a month for about 0.5 hour each time to ensure that all components remain in good working order.


Note: Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. is a specialized manufacturer engaged in the R&D, production, and sales of medical centrifuge products. Our current product lineup includes high-speed centrifuges, low-speed centrifuges, large-capacity centrifuges, and refrigerated centrifuges, and we also offer custom-made centrifuges tailored to specific customer requirements. Our products are widely used in research and production institutions across various fields, including agricultural science, bioengineering, food processing, chemical industry, pharmaceuticals, clinical medicine, blood banks and blood stations, inspection and quarantine, disease control and prevention, and environmental protection.