Besides the rotor, what other aspects should be taken into account when using a desktop centrifuge?
Release time:
2020-06-02
Overview:
Desktop centrifuge It utilizes the powerful centrifugal force generated by the high-speed rotation of a centrifuge rotor to accelerate the sedimentation rate of particles in a liquid, thereby separating substances in a sample that have different sedimentation coefficients and buoyant densities. It has been widely applied in fields such as biomedicine, petrochemicals, agriculture, and food hygiene.
The development trends of benchtop centrifuges are as follows: 1. Centrifuges are adopting multi-functionality, automation, and microcomputer control; 2. AC variable-frequency speed regulation will gradually replace DC speed regulation in centrifuges; 3. Centrifuge rotors are being improved and their variety is increasing; 4. Centrifuges are paying greater attention to safety, environmental protection, and compliance with international standards.
The above outlines the new trends in desktop centrifuge development. Customers purchasing desktop centrifuges should consider all factors comprehensively and make their selection accordingly.
The main difference between centrifuge models lies in their performance. Benchtop centrifuges equipped with refrigeration are significantly more expensive than standard ones. Some centrifuges also feature heating functions, and the more sophisticated the control program, the higher the price of the benchtop centrifuge. A major distinction is found in the equipment configuration—sometimes the cost of accessories can even exceed the price of the main unit itself. When selecting a benchtop centrifuge, it’s important to note that, in addition to the main unit itself, you’ll need to consider the rotor you choose, along with the necessary centrifuge tubes, tube holders, specialized centrifuge bottles, or blood bags. Only when you add up all these components can you determine the full cost of a complete benchtop centrifuge system. Beyond these factors, the visual difference between a benchtop centrifuge and a conventional centrifuge lies in their size—of course, the benchtop model is designed to be smaller in volume. This compact design ensures that the benchtop centrifuge won’t take up too much laboratory space.
A desktop centrifuge is primarily used to extract the desired components from liquid mixtures. By spinning at high speed, it takes advantage of the differences in density among various substances: denser liquids settle at the bottom, while less dense liquids float on top, thus separating the liquid into distinct layers and allowing us to isolate the pure substances we need. It finds applications in many fields, including pharmaceuticals and chemical engineering.
Issues to watch out for:
1. The volume of solution loaded into the centrifuge tubes must be appropriate. For stainless steel tubes without caps, fill them only to 2/3 capacity; for plastic tubes, you can fill them up to the shoulder. The tube caps must be tightly sealed—leakage is absolutely not allowed. Empty tubes may deform during centrifugation. When using organic solvents with plastic tubes, ensure compliance with applicable regulations.
2. Centrifuge tubes must be placed in pairs and strictly balanced, with a deviation no greater than 0.1 g.
3. Do not allow the rotor to spin idly without a sample. Always use the handle when placing or removing the rotor to prevent it from slipping and falling. Place the rotor gently and ensure it is securely locked in place. When unscrewing the handle, hold the handle with your hand and rotate only the rotor itself. The rotor cover must be tightly sealed; centrifugation is prohibited without the cover properly in place.
4. Do not open the machine cover during centrifugation, and do not lean against or hold onto the centrifuge. If you hear any unusual noises or notice vibrations, immediately stop the machine.
5. After using the rotor, it must be promptly removed from the rotor chamber, dried thoroughly, and then carefully cleaned inside the tube holes using a hole-cleaning rod. If any solution has spilled, it must be thoroughly cleaned up. Wipe away any condensation inside the rotor chamber, open the door to allow the chamber to air-dry.
6. Use of the centrifuge requires a reservation and must be logged after use.
The differential centrifugation method for desktop centrifuges involves gradually varying the centrifugal speed or alternately performing high-speed and low-speed centrifugations. By selecting centrifugal forces of different magnitudes, this method allows tiny particles with varying weights and sedimentation coefficients to settle out of the suspension in successive batches, thereby achieving separation. The differential centrifugation method is particularly suitable when there are significant differences in weight—or, more precisely, in sedimentation coefficients—among the particles in the suspension. The greater the difference in weight or sedimentation coefficient, the better the separation effect. For particles with small differences in weight or sedimentation coefficients, the differential centrifugation method struggles to achieve satisfactory separation results. Since the differential centrifugation method relies on either obtaining a precipitate or collecting the supernatant after removing larger particles, it is also known as differential sedimentation.
When a tabletop centrifuge rotor spins at high speed, it generates centrifugal stress internally. Once this centrifugal stress exceeds a certain threshold, the rotor may fail prematurely or even explode. Therefore, users must pay close attention to the proper use and careful storage of rotors. After each use, the rotor should be promptly cleaned and never subjected to impacts or scratches. If the rotor is accidentally dropped, it must undergo an X-ray examination to ensure there is no internal damage before it can be used again. During operation, rotors that have exceeded their service life should be promptly discontinued. Typically, manufacturers specify both the total operating hours and the maximum number of centrifugation cycles for each rotor; whichever of these two indicators is reached first indicates that the rotor has reached the end of its service life. To extend the lifespan of a tabletop centrifuge rotor, users must strictly follow the manufacturer’s instructions.