Centrifuges are widely used in pharmaceutical manufacturing. How should pharmaceutical companies make a rational selection?
Release time:
2020-01-08
Overview:
Centrifuge As a solid-liquid separation device, it is one of the key pieces of equipment used by pharmaceutical manufacturers and is widely applied in the pharmaceutical industry. For example, in biopharmaceutical production, it is often necessary to separate cells (microbial biomass), proteins, enzymes, DNA, peptides, and other substances from fermentation broth—this process relies on the centrifugal force generated by the high-speed rotation of a centrifuge. Similarly, conventional blood separation and purification of drug components also commonly employ centrifuges.
Centrifuges play a crucial role in the pharmaceutical manufacturing process. However, the pharmaceutical industry must adhere to the GMP requirements outlined in the “Good Manufacturing Practice for Pharmaceutical Products.” These requirements also encompass specific stipulations for pharmaceutical machinery, equipment, and instruments. Therefore, centrifuges must be designed, selected, and installed in strict compliance with pharmaceutical production standards.
“In pharmaceutical manufacturing, centrifuges that are appropriately selected, easy to clean and disinfect, convenient for production operations, capable of preventing errors, and effective in reducing contamination are increasingly favored by the market,” said an industry insider. The author has learned that, in recent years, Chinese centrifuge manufacturers have been steadily striving forward to further promote the development of centrifuge technology in China and enable centrifuges to play an even greater role in pharmaceutical production. As a result, centrifuge technology in China has made rapid and remarkable progress.
However, in the actual production process, how should pharmaceutical manufacturers correctly select centrifuges, and how should they properly determine parameters and rotor specifications? How can pharmaceutical manufacturers choose the right sterilization method, and what steps can they take to improve centrifugation performance? To further promote the widespread application of centrifuges in the pharmaceutical industry, advance the development of the pharmaceutical sector, and benefit the people, Shanghai Luxiangyi— as a centrifuge manufacturer—recently shared its expertise in centrifuge technology with everyone.
I. How Pharmaceutical Manufacturers Can Correctly Select the Main Parameters of Centrifuges
Centrifuges are indispensable equipment in biopharmaceutical production. So, how can you choose a centrifuge correctly? According to technical staff from Shanghai Luxiangyi, the following points should be considered when selecting a centrifuge:
1. Selection of the maximum relative centrifugal force
Pharmaceutical manufacturers select the maximum relative centrifugal force required for separation based on different process requirements. The manufacturer will indicate the maximum relative centrifugal force of the centrifuge in the instruction manual or promotional materials. This maximum relative centrifugal force is denoted by RCF, also known as the separation factor, and plays a dominant role during the centrifugation process. When selecting a centrifuge, pharmaceutical manufacturers sometimes pay attention only to the rotational speed while neglecting the separation factor—a common misconception. This is because the separation factor is related to both the square of the rotational speed and the radius of rotation, as shown in the following formula:

RCF: The maximum relative centrifugal force is expressed in multiples of the acceleration due to gravity, denoted as ×g.
n: The maximum rotational speed of the rotor, in units of r/min;
r: The maximum centrifugal radius of the rotor, in centimeters;
g: Acceleration due to gravity.
Consider the following figure to calculate RCF:

Shanghai Luxiangyi Centrifuge Figure 1
2. Selection of Rotational Speed
After selecting the RCF, you can correspondingly select the rotor speed of the centrifuge.
3. Selection of Capacity
Select the rotor capacity that matches both the RCF and the required rotational speed. The manufacturer will provide these parameters on its website, in brochures, or in the instruction manual. Centrifuges can be equipped with multiple rotors of different capacities, allowing pharmaceutical manufacturers to choose according to their specific needs. Only by satisfying all three of the above criteria can the centrifuge meet the pharmaceutical manufacturer’s operational requirements.
4. Selection of refrigerated or non-refrigerated centrifuges
People in pharmaceutical companies often ask: “If I don’t need to freeze my samples, why should I choose a refrigerated centrifuge?” The reason is that during high-speed rotation, the rotor and the air within the chamber generate significant heat due to friction. Additionally, the motor itself produces heat when operating at high speeds. If you don’t opt for a refrigerated centrifuge, the temperature inside the chamber will rise rapidly after the centrifuge starts running—potentially reaching as high as 80 degrees Celsius. This elevated temperature can compromise the activity of the desired active ingredients in your samples and also shorten the lifespan of the centrifuge. Therefore, for centrifuges used in pharmaceutical applications, it’s essential to choose a refrigerated model.
II. How Pharmaceutical Manufacturers Can Correctly Select Rotors
The rotor is a crucial component of a centrifuge. There are several types of rotors, including angle rotors, horizontal rotors, and continuous-flow rotors. Angle rotors have a small centrifugal radius and high rotational speed, but their capacity is relatively smaller than that of horizontal rotors, making them ideal for separating and extracting fine particles. Horizontal rotors are better suited for separating larger particles; although their rotational speed is lower than that of angle rotors, they offer a significantly larger capacity. Continuous-flow rotors enable quantitative and continuous production, but they are not suitable when the solid content is high. This is because loading and unloading continuous-flow rotors and retrieving the sediment can be cumbersome, time-consuming, and require operators to be highly experienced and possess considerable arm strength. If the solid content exceeds 10%, it is not recommended to choose this type of rotor.
Example 1: For protein separation, choose a high-speed refrigerated centrifuge.
Rotor: Angle rotor, RCF = 14,400 × g, Speed = 9,000 rpm, Capacity = 500 ml × 6
Example 2: When extracting enzymes from animal viscera, choose a large-capacity refrigerated centrifuge.
Rotor: Horizontal rotor, RCF = 6260 × g, Speed = 4000 rpm, Capacity = 2000 ml × 6
Example 3: When collecting bacterial debris by centrifugation, choose a high-speed refrigerated centrifuge.
Rotor: Continuous-flow rotor, RCF = 28,100 × g, Speed = 16,000 rpm, Capacity = 2,000 ml × 6
Example 4: For centrifugal selection of E. coli, use a high-speed refrigerated centrifuge.
Rotor: Angle rotor, RCF = 11,600 × g, Speed = 7,000 rpm, Capacity = 1,000 ml × 6
III. How pharmaceutical manufacturers can correctly select a sterilization method:
The production of biopharmaceuticals, especially vaccines, places extremely stringent requirements on process hygiene—particularly those stipulated in GMP certification. In centrifuges, all components that come into contact with the material, except for the rotor of continuous-flow centrifuges, are in contact with centrifuge bottles. During autoclaving, it is essential to use the moist-heat sterilization method at a temperature of 121°C. The caps of the centrifuge bottles must be opened during sterilization. If the caps remain closed during sterilization, the centrifuge bottles (made of PP, PE, or PC) will generate gas under high temperature and pressure, leading to deformation and rendering them unusable.
4. How can pharmaceutical manufacturers achieve optimal centrifugation performance?
After the centrifuge is purchased, the manufacturer will typically come to your facility for installation and commissioning, as well as to train your operators. At this stage, some pharmaceutical companies may not yet have actual materials requiring centrifugation—instead, they’ll just perform a basic setup and initial debugging of the equipment. Once it’s time to start centrifuging a batch, first carefully read the instruction manual to reinforce your understanding. Then, select the appropriate centrifugal force and corresponding rotor speed, and determine the correct centrifugation duration. The choice of centrifugation time should align with the pharmaceutical company’s specific process requirements: too long a duration reduces efficiency, while too short a duration may compromise the separation effect. Therefore, it’s crucial to choose a centrifugation time that’s well-suited to the actual material being processed. Next, set the temperature; unless otherwise specified, pharmaceutical companies usually opt for 4°C. Also, configure the acceleration and deceleration settings. During the centrifuge installation and commissioning, it’s generally recommended to set the acceleration to level 4 and the deceleration to level 5. However, certain specialized pharmaceutical processes may require deceleration to the slowest setting. In some cases, the procedure might even call for a specially designed deceleration curve. All such requirements need to be clearly communicated at the time of purchase. For example, when separating calf serum, if the standard deceleration rate is used, the residual braking force at the end of the cycle could cause the previously separated plasma and red blood cells to become cloudy again. To avoid this issue, a special centrifugation program with a gradual, slow deceleration must be designed. Only in this way can you achieve the optimal separation results.
V. How Pharmaceutical Manufacturers Should Draft the Technical Requirements Section of the URS Document
The URS document is a standardized document prepared by pharmaceutical manufacturers to outline procurement requirements. It covers a wide range of content, including commercial and technical aspects. The key technical sections related to centrifuges, along with their corresponding writing formats, are as follows:
1) The maximum relative centrifugal force shall be no less than xxxx*g.
2) Single centrifugation capacity ≥ xxxx L.
3) It can effectively maintain the temperature inside the centrifugal chamber at xx–xx℃, with temperature fluctuations no greater than xx℃. Within half an hour of turning on the cooling system, the temperature can be brought down to xx–xx℃.
4) The installation, commissioning, operation, and maintenance of the machine must all comply with environmental protection requirements, and the operating noise level must be ≤ xx dB(A).
5) The centrifuge is equipped with an interlock control function that links the opening of the door cover to the equipment’s operating status, ensuring that the centrifuge door cover cannot be opened during operation.
6) The equipment is equipped with an emergency stop control switch, which should be located in an easily accessible position.
7) Alarm devices such as overtemperature protection, overspeed protection, door/cover protection, and imbalance protection shall be provided.
8) The material of the centrifuge cup should possess properties such as wear resistance, resistance to high-temperature sterilization, and tolerance to commonly used solvents.
9) The equipment’s parameter settings and operation shall be controlled via the display screen, which allows for setting of rotational speed or centrifugal force. The display screen features a Chinese interface and should be able to show parameters such as rotational speed, centrifugal force, temperature, and running time during centrifugation.
10) Each electrical circuit shall be clearly labeled and positioned conspicuously. Electrical control components shall have complete nameplates and markings to facilitate maintenance and inspection.
11) Provide a list of spare parts and their prices for the next three years.
VI. Safety Precautions for Centrifuge Use
When a centrifuge is operating at high speed, even minor operational errors can cause serious damage to the machine. Understanding safety precautions, precautions, and proper operating procedures can help prevent injuries to operators and ensure that the centrifuge’s performance remains unaffected.
1) Pre-experiment inspection items include: checking whether the rotor and other accessories are in good condition, ensuring that the rotor is properly seated on the spindle, verifying that the center nut or pressure head on the rotor body is securely tightened, and confirming that the centrifuge bottles/centrifuge tubes are free of deformation or cracks.
2) Weigh and balance the centrifuge bottles/centrifuge tubes on a balance, adjusting the amount of centrifuged material to be loaded so that the weights of the centrifuge bottles/centrifuge tubes containing the centrifuged material are as equal as possible. Do not exceed the maximum allowable imbalance, typically 2 grams. Then, symmetrically place the centrifuge bottles/centrifuge tubes into the rotor.
3) Before starting the centrifuge, make sure the centrifuge chamber has been thoroughly cleaned and is free of any debris.
4) While the machine is running, do not open the centrifuge door cover, and definitely do not move the centrifuge. Do not place any items on the door cover to prevent damage to those items if the door cover opens automatically.
5) After centrifugation is complete, the speed will gradually decrease. Only after the rotor has come to a complete stop can you press the “Stop” button to open the lid and access the centrifuge door to remove the centrifuge bottles or tubes. When removing the centrifuge bottles or tubes, be careful not to disturb the settled particles by shaking, which could cause them to resuspend and lead to turbidity.
6) When using a centrifuge, if the screen displays a fault alarm during operation, you should immediately stop the machine or disconnect the power supply. If the centrifuge vibrates excessively, it indicates that the instrument is unbalanced. If you hear the sound of a centrifuge bottle or tube breaking inside the instrument, you should immediately cease use and inspect the equipment. Before using it again, be sure to thoroughly remove all fragments.
7) After use, be sure to clean it thoroughly and disconnect the power supply.
7. Maintenance of the Centrifuge
1) After each use of the instrument, first remove the rotor, then clean and dry the centrifuge chamber with a cotton cloth. Once the rotor is removed from the centrifuge chamber, it should be promptly washed with a neutral detergent, dried thoroughly with a dry cotton cloth, and stored in a dry, well-ventilated area. Never clean the rotor with non-neutral detergents to prevent chemical corrosion.
2) If the centrifuge is not used for an extended period, clean the interior cavity of the centrifuge and place a desiccant inside to absorb moisture, preventing the drive shaft from rusting. Apply a small amount of grease to the motor shaft for protection.
3) The surface of the control panel must be kept clean. The control panel should only be cleaned with a dry, soft cloth and must never come into contact with water to prevent water from entering the control circuit board and damaging it.
4) Do not allow any impact or damage to the conical surface of the drive shaft. Before installing the rotor, use a soft cloth to thoroughly clean both the drive shaft and the rotor’s conical bore, and apply a small amount of medical petroleum jelly or other lubricant.
5) The rotor, centrifuge tubes or hanging cups, and other accessories must not be placed on the centrifuge door cover or the machine head to prevent scratches and damage. The exterior surfaces of the door cover and the machine casing should be cleaned only with a soft cloth and a neutral detergent to avoid peeling of the paint layer. The use of highly corrosive disinfectants (such as 84 Disinfectant) for sterilizing the centrifuge is strictly prohibited.
The above are some insights and experiences gained during the communication process between Shanghai Luxiangyi and pharmaceutical manufacturers, for your reference.